Deep Well Plates: Volumes, Dimensions and Bottom Shapes
By Arctic White Team. Part of the Learning Center.
The short answer
A deep well plate is a storage plate on the standard ANSI/SLAS (SBS) footprint, 127.76 x 85.47 mm, with taller wells than an assay plate. In 96 well format the wells hold from about 0.5 mL to 2.2 mL each. Pick the volume first, then the bottom shape: pyramid and conical bottoms concentrate a small remaining volume in one place, round (U) bottoms suit mixing and resuspension, and the volume that matters day to day is the working volume, not the maximum.
One footprint, four depths
Every plate below sits on the same footprint, so a liquid handler, a centrifuge rotor and a sealing mat see the same plate outline. What changes is height and well depth. From the product pages, checked 2026-11-20:
| SKU | Well volume | Well / bottom | Plate height | Well depth |
|---|---|---|---|---|
| AWLS-X10005 | 1.6 mL working (1.7 mL max) | Round / round | Not listed | Not listed |
| AWLS-S30009 | 2 mL per well | Square / pyramid | 44.04 mm | 39.22 mm |
| AWLS-219009 | 2 mL working (2.2 mL max) | Square / pyramid | 44.4 mm | 39.9 mm |
| AWLS-X11005 | 2.0 mL working (2.1 mL max) | Round / round (U) | 45.30 mm | 42.60 mm |
All four: 96 wells, polypropylene, ANSI/SLAS (SBS) footprint. Browse the range in 96 well storage and deep well plates.
Maximum, working and capped volume are three different numbers
Maximum volume is what a well holds to the brim. Working volume is what you can pipette, shake and freeze without spilling. Capped volume is what fits once a sealing mat sits on top. On AWLS-X10005 those numbers are 1.7 mL maximum, 1.6 mL working and 1.5 mL with a cap. On AWLS-X11005 they are 2.1 mL maximum, 2.0 mL working and 1.9 mL with a mat lid. Compare the same number across plates, and plan your fill against the working volume.
Bottom shapes: what each one is for
Pyramid bottom (square wells): the well floor slopes to a central point, so the last of a sample collects in one place for a pipette tip or a needle. Our square well geometry is worked through in Why square wells?
Conical (V) bottom: a pointed round well. It concentrates small volumes the same way, and suits pelleting in a centrifuge.
Round (U) bottom: no corners and a curved floor. Samples resuspend without sitting in corners, which is why round wells are the default for mixing and cell or bacterial culture. See Why round wells?
Flat bottom: the well floor is level, which suits optical reads from below. Flat bottoms are assay plate territory more than deep well territory.
Will it fit your instrument?
The ANSI/SLAS standards fix the footprint (127.76 x 85.47 mm), the 9.0 mm well pitch in 96 well format, and the position of well A1 (14.38 mm from the left edge, 11.24 mm from the top on AWLS-S30009). Height is not fixed by the standard: it runs from about 44 mm to 45.3 mm across the 2 mL plates above. Check plate height against your handler's gripper and your centrifuge buckets before you switch plates, and check the sealing mat against the well shape: a square well takes a square plug mat, a round well takes a round one.
Temperature and centrifuge ratings differ by plate
AWLS-X11005 is rated from -80 °C to 120 °C, is autoclavable at 121 °C, and carries a maximum centrifugation rating of 4,000 RCF. AWLS-S30009 is rated from -86 °C. AWLS-219009 is rated for storage down to -86 °C. Read the rating on the plate you are buying rather than assuming the range shares one.
FAQ
What is the standard size of a 96 deep well plate?
The footprint is the ANSI/SLAS (SBS) standard, 127.76 x 85.47 mm, with a 9.0 mm well pitch. Height varies by volume: the 2 mL plates on this page run from 44.04 mm to 45.30 mm tall.
What is the difference between maximum volume and working volume?
Maximum volume is what the well holds to the brim. Working volume is what you can pipette, shake and freeze without spilling. On AWLS-X10005 the figures are 1.7 mL maximum and 1.6 mL working, and 1.5 mL once a cap mat is on.
Which bottom shape recovers the most sample?
Pyramid and conical (V) bottoms concentrate the last of a sample in one place, so a tip can reach it. Round (U) bottoms leave no corners, which suits mixing and resuspension. The right choice depends on whether recovery or mixing matters more in your method.
Can deep well plates be frozen?
The plates on this page are rated for storage down to -80 °C or -86 °C, depending on the SKU. Check the rating on the product page, and seal with a mat rated for the same temperature before freezing.
Sources
Product pages for AWLS-X10005, AWLS-S30009, AWLS-219009 and AWLS-X11005 on arcticsci.com, checked 2026-11-20. Footprint and pitch are the ANSI/SLAS (SBS) standard values those pages list.